CVE-2025-15653 — Dräger Zeus IE Anesthesia Workstation USB Interface Privilege Escalation
Dräger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical acces…
Published: 2026-06-02 · CVSS: 7.0 · CVSS 4.0
Dräger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dräger Service Connect.
Source: CVE Program · Revision 8b16575c39f1e1810c5e863010da073d2d3285e7