CVE-2024-48967 — Life2000 ventilator and Service PC lack sufficient audit logging capabilities
The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with a…
Published: 2024-11-14 · CVSS: 10.0 · CVSS 3.1
The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance.
Source: CVE Program · Revision 8b16575c39f1e1810c5e863010da073d2d3285e7