CVE-2022-38100 — Contec Health CMS8000
The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP …
Published: 2022-09-13 · CVSS: 7.5 · CVSS 3.1
The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network.
Source: CVE Program · Revision 8b16575c39f1e1810c5e863010da073d2d3285e7