CVE-2019-6540 — Medtronic Conexus Radio Frequency Telemetry Protocol Cleartext Transmission of Sensitive Information
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Cla…
Published: 2019-03-26 · CVSS: 6.5 · CVSS 3.1
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement encryption. An attacker with adjacent short-range access to a target product can listen to communications, including the transmission of sensitive data.
Source: CVE Program · Revision 8b16575c39f1e1810c5e863010da073d2d3285e7