datAInsights CVE Current

CVE-2019-25721 — Dräger Infinity M300 VG2.3.1 Network-Based Denial of Service

Dräger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attacke…

Published: 2026-06-02 · CVSS: 7.1 · CVSS 4.0

Dräger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality.

Original CVE JSON · STIX 2.1

Source: CVE Program · Revision 8b16575c39f1e1810c5e863010da073d2d3285e7