datAInsights CVE Current

CVE-2018-8870 — Medtronic MyCareLink Patient Monitor Use of Hard-coded Password

Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains a hard-coded operating system password. An attacker with physical access can remove the case of the devi…

Published: 2018-07-02 · CVSS: 6.4 · CVSS 3.1

Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains a hard-coded operating system password. An attacker with physical access can remove the case of the device, connect to the debug port, and use the password to gain privileged access to the operating system.

Original CVE JSON · STIX 2.1

Source: CVE Program · Revision 8b16575c39f1e1810c5e863010da073d2d3285e7