datAInsights CVE Current

CVE-2018-14781 — Medtronic MiniMed MMT-500/MMT-503 Remote Controllers Authentication Bypass by Capture-replay

Medtronic MiniMed MMT devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-re…

Published: 2018-08-13 · CVSS: 5.3 · CVSS 3.1

Medtronic MiniMed MMT devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.

Original CVE JSON · STIX 2.1

Source: CVE Program · Revision 8b16575c39f1e1810c5e863010da073d2d3285e7