CVE-2017-6018
An open redirect issue was discovered in B. Braun Medical SpaceCom module, which is integrated into the SpaceStation docking station: SpaceStation with SpaceCom module (integrat…
Published: 2017-06-30 · CVSS: Unscored
An open redirect issue was discovered in B. Braun Medical SpaceCom module, which is integrated into the SpaceStation docking station: SpaceStation with SpaceCom module (integrated as part number 8713142U), software versions prior to Version 012U000040, and SpaceStation (part number 8713140U) with installed SpaceCom module (part number 8713160U), software versions prior to Version 012U000040. The web server of the affected product accepts untrusted input which could allow attackers to redirect the request to an unintended URL contained within untrusted input.
Source: CVE Program · Revision 8b16575c39f1e1810c5e863010da073d2d3285e7