datAInsights CVE Current

CVE-2017-5149

An issue was discovered in St. Jude Medical Merlin@home, versions prior to Version 8.2.2 (RF models: EX1150; Inductive models: EX1100; and Inductive models: EX1100 with MerlinOn…

Published: 2017-02-13 · CVSS: Unscored

An issue was discovered in St. Jude Medical Merlin@home, versions prior to Version 8.2.2 (RF models: EX1150; Inductive models: EX1100; and Inductive models: EX1100 with MerlinOnDemand capability). The identities of the endpoints for the communication channel between the transmitter and St. Jude Medical's web site, Merlin.net, are not verified. This may allow a man-in-the-middle attacker to access or influence communications between the identified endpoints.

Original CVE JSON · STIX 2.1

Source: CVE Program · Revision 8b16575c39f1e1810c5e863010da073d2d3285e7