CVE-2015-3459
The communication module on the Hospira LifeCare PCA Infusion System before 7.0 does not require authentication for root TELNET sessions, which allows remote attackers to modify…
Published: 2015-04-29 · CVSS: Unscored
The communication module on the Hospira LifeCare PCA Infusion System before 7.0 does not require authentication for root TELNET sessions, which allows remote attackers to modify the pump configuration via unspecified commands.
Source: CVE Program · Revision 8b16575c39f1e1810c5e863010da073d2d3285e7